Calibration and Calibration Verification

Each section must perform and document the number and concentration of calibrators used for calibration procedures following the manufacturer’s instructions or use the criteria established when establishing or verifying performance specifications according to CAP/CMS standards. Established calibration procedures must be performed and documented at the established frequency, but no less than once every 6 months, and each time any of the following occur:

Quality Control

Each clinical section must have a clearly defined and documented Quality Control policy that meets CAP/CMS standards, including the establishment of number, type, and frequency of testing control materials. For each test system, the section is responsible for having control procedures that monitor the accuracy and precision of the complete analytical process. The control procedures must:

Results of control materials must meet laboratory and, as applicable, manufacturer’s criteria for acceptability before reporting patient test results. The laboratory must document all control procedures performed.

The Quality Control records are maintained in each laboratory and reviewed regularly by each laboratory supervisor or technical quality control coordinator. The Quality Control records are maintained for a minimum of two (2) years. The Quality Control programs have clearly defined goals for monitoring analytic performance, procedures, policies, tolerance limits, corrective action and related information in accord with CAP standards.

When neither calibration nor control materials are available for a particular test, each section must establish procedures to verify the reliability of patient test results.

As part of the department-wide Quality Control program, interim CAP self-inspections are performed, in which deficiencies as well as corrective actions taken are documented.

Special Stains and In-Situ Hybridization (ISH) Quality Control

For special stains, including histochemical stains, and studies using immunologic and ISH methodology, positive and negative controls are verified and recorded as acceptable prior to or concurrent with the reporting of patient results and records maintained.

For immunofluorescence microscopy, appropriate positive and negative controls must be performed.

For Immunohistochemistry antibodies, a positive and negative control must be performed for each antibody. Immunohistochemical tests using polymer-based detection systems (biotin-free) are sufficiently free of background reactivity to obviate the need for a negative reagent control and such controls may be omitted at the discretion of the laboratory director following appropriate validation. 

In situ hybridization controls depend on the specific assay, signal pattern present, and sample type. For example, assays designed to detect deletions may use internal controls that include both the probe of interest and a control locus probe, both of which map to the same chromosome. In this situation, there are two internal controls, the signal for the probe of interest on the normal homolog and the control locus signals on both the normal and deleted homolog. For a dual fusion assay, the probe signals on each of the normal homologs function as internal controls. If a probe is used that does not produce an internal control signal (e.g. a Y chromosome probe in a female), another sample that is known to have the probe target must be run in parallel as an external control with the patient sample. In addition, many ISH assays use an external control(s). For FDA-cleared or approved ISH assays, laboratories must follow manufacturer's instructions for quality control at minimum.

Results of controls must be recorded, either in internal laboratory records, or in the patient report. A statement in the report such as, "All controls show appropriate reactivity" is sufficient.