Each laboratory section defines the type of water necessary for each of its procedures as required by the test manufacturer or published guidelines. The current edition of the Clinical Laboratory Standards Institute (CLSI) Guideline GP40-A4-AMD defines the following grades of water: Clinical Laboratory Reagent Water (CLRW), suitable for most laboratory procedures; Special Reagent Water (SRW), defined by a laboratory for procedures that need different specifications than CLRW; Instrument Feed Water, specified by IVD manufacturers as suitable for use with their measurement systems; and Commercially Bottled Purified Water that may be suitable for certain laboratory procedures. CLRW is similar to the Type I reagent water defined in earlier editions of this guideline. CLRW is not required if the laboratory is able to document reliable results with an alternate grade of water.
The following specification for CLRW is adapted from this guideline and should be met at time of in-house production:
The technical staff of each area confirms water resistance with each use and troubleshoots the purification equipment if necessary (e.g., change filters, change cartridges, place a service call). In addition, resistivity is checked after maintenance performed by the Neu Ion or Millipore (current vendors) representative, or documented from the readout on NeuIon or Millipore Systems.
The quality (specifications) of the laboratory's water, whether prepared in-house or purchased, must be checked and documented at least annually. The frequency and extent of checking may vary, according to the quality of source water and specific laboratory needs. Corrective action must be documented if water does not meet acceptability criteria.
Reagent grade water (type I and II) is monitored for bacteria annually. The Clinical Manager or designated technical staff member performs the bacterial monitoring according to the instructions below. After the cultures are analyzed, results are reviewed by his/her designated supervisor and distributed to each section.
Each section may opt-out and perform individual microbial checks, and will be responsible for corrective actions and retaining all records.
| Type I | Type II | Type III |
---|---|---|---|
Maximum microbial content (CFU/ml) | 10 | 1000 | n/a |
Minimum resistivity (megohm-cm) | 10 (in-line) | 1.0 | 0.1 |
Maximum silicate content (mg/L SiO2) | 0.05 | 0.1 | 1.0 |
Particulate matter | 0.22 filter | n/a | n/a |
Copies of the completed Bi-Yearly QC worksheet are distributed to each section.
Each section must document procedures for handling and cleaning glassware, where this is done. Documentation must include methods for testing for detergent removal.
In order to detect detergent residues resulting from improper rinsing, a pH meter or pH paper qualiyt control (QC) method can be used:
a. Rinse a small clean beaker by filling and emptying 3 times with source water.
b. Fill a fourth time and measure pH using a pH meter or pH paper. Record the pH as source water pH.
c. Take a piece of newly-cleaned glassware in order to test, fill about 10% full with source water. Use more water if necessary to get enough water to be able to sufficiently immerse the pH meter electrode or pH paper.
d. Agitate water in glassware to extract residues from all possible surfaces.
e. Take pH reading with pH meter or paper and record as glassware pH
f. Any significant increase in pH indicates alkaline detergent residue. A significant change is 0.2 or more pH units measuring to 0.1 pH units of sensitivity. A result of less than 0.2 pH unit change indicates properly rinsed glassware.
Note regarding pH paper: detergents and surface active agents can interfere with some pH paper by causing a decrease of several pH units in reading. Please test any pH paper you intend to use with these detergents to determine if there is any interference before implementing pH paper method.
Clinical and Laboratory Standards Institute. Preparation and Testing of Reagent Water in the Clinical Laboratory; Approved Guideline. 4th ed. CLSI document GP40-A4-AMD. Clinical and Laboratory Standards Institute, Wayne, PA, 2012.
Stewart BM. The production of high-purity water in the clinical laboratory. Lab Med. 2003;31:605-611.
LP has evaluated its source water to determine if a high concentration of silicates is present. Results are within normal limits (< 0.05 ppm) and may be viewed by contacting the Clinical Manager.